Skip to main content

Table 4 Mean differences and 95% limits of agreement between measurements obtained with μCT and CBCT images at different tube voltage settings

From: Evaluation of trabecular bone microstructure and cortical morphology using cone-beam and micro-CT images: impact of tube voltage setting

 

Mean difference ± standard deviation

Limits of agreement (95%)

p-value

Lower

Upper

Trabecular measurements

BV/TV (%)

 μCT-CBCT80

−2.16 ± 11.51

−24.72

20.40

0.3111

 μCT-CBCT85

−2.83 ± 12.13

−26.60

20.40

0.2110

 μCT-CBCT90

−2.33 ± 9.91

−21.75

17.09

0.2272

Tb.Th (mm)

 μCT-CBCT80

−0.41 ± 0.05

−0.93

0.11

< 0.0001*

 μCT-CBCT85

−0.41 ± 0.05

−0.94

0.11

< 0.0001*

 μCT-CBCT90

−0.43 ± 0.05

−0.93

0.07

< 0.0001*

Tb.Sp (mm)

 μCT-CBCT80

−0.40 ± 0.12

−1.65

0.83

< 0.0001*

 μCT-CBCT85

−0.51 ± 0.11

−1.66

0.64

< 0.0001*

 μCT-CBCT90

−0.43 ± 0.11

−1.65

0.69

< 0.0001*

Cortical measurements

Ct.Ar (mm2)

 μCT-CBCT80

−5.26 ± 111.23

−1199

1188

0.9626

 μCT-CBCT85

−38.73 ± 111.22

−1232.74

1155.27

0.7302

 μCT-CBCT90

−23.63 ± 113.98

−1247.33

1200.06

0.8372

Ct.Th (mm)

 μCT-CBCT80

−0.42 ± 0.07

−1.16

0.31

< 0.0001*

 μCT-CBCT85

−0.45 ± 0.08

−1.26

0.35

< 0.0001*

 μCT-CBCT90

−0.44 ± 0.07

−1.21

0.33

< 0.0001*

  1. μCT micro-CT, CBCT cone beam computer tomography, SD standard deviation, BV/TV bone volume fraction, Tb.Th trabecular thickness, Tb.Sp trabecular separation, Ct.Ar cortical area, Ct.Th cortical thickness
  2. *Statistical significance (p < 0.01)