Skip to main content

Table 1 Means and SD of different EALs at various temperatures

From: Evaluation of heated sodium hypochlorite’s effect on the accuracy of contemporary electronic apex locators: an in vitro study

 

19.4 °C (± 1.5 °C)

36o

70o

 

Mean ± SD (Min/Max)

Mean ± SD (Min/Max)

Mean ± SD (Min/Max)

2p

Root ZX Mini

-0.10 ± 0.2000 (-0.6/0.32)

-0.18 ± 0.25 (-0.65/0.45)

-0.16 ± 0.19 (-0.55/0.28)

0.265

Propex Pixi

-0.11 ± 0.29 (-0.57/0.54)

-0.20 ± 0.23 (-0.52/0.31)

-0.17 ± 0.24 (-0.67/0.25)

0.370

Raypex 6

-0.11 ± 0.25 (-0.44/0.67)

-0.24 ± 0.23 (-0.58/0.30)

-0.15 ± 0.26 (-0.52/0.46)

0.113

Ai-Pex

-0.10 ± 0.23 (-0.55/0.43)

-0.10 ± 0.22 (-0.42/0.40)

-0.09 ± 0.20 (-0.44/0.20)

0.980

1p

0.995

0.272

0.682

 
  1. 1Oneway ANOVA Test 2Repeated Measures Analysis of Variance